Model |
Chroma Meter CS-100A |
Type |
SLR spot colorimeter for measuring light-source and surface luminance and chromaticity |
Measuring angle |
1˚ |
Optical system |
85mm f/2.8 lens; SLR viewing system; flare factor less than 1.5% |
Angle of view |
9˚ with 1˚ measurement area indication |
Focusing distance |
1014mm to infinity |
Receptors |
3 silicon photocells filtered to detect primary stimulus values for red, green and blue light |
Spectral response |
Closely matches CIE 1931 Standard Observer curves |
Response time |
FAST: Sampling time: 0.1s, Time to display: 0.8 to 1.0s; SLOW: Sampling time: 0.4s, Time to display: 1.4 to 1.6s |
Luminance units |
cd/m2 or fL |
Measuring range |
FAST 0.01 to 299,000cd/m2
SLOW 0.01 to 49,900cd/m2 |
Accuracy |
Luminance(Y): ±2% of reading ±1digit
Chromaticity (x, y):
FAST Y 100cd/m2 or above: ±0.001; 48.1 to 99cd/m2 : ±0.002; below 48.1cd/m2: below measurement range
SLOW Y 25.0cd/m2 or above: ±0.001; 12.0 to 24.9cd/m2 : ±0.002; below 12.0cd/m2: below measurement range
(Measurement subject: Illuminant A) |
Target value |
1; set by measurement or numerical input |
Measurement modes |
Absolute color: Yxy; color difference: Δ(Yxy) |
Data communication |
RS-232C |
External control |
Measurement can be started by external device connected to data output terminal |
Power source |
One 9V battery, Power can be supplied via data output terminal |
Operating temperature/humidity range |
0 to 40℃, relative humidity 85% or less(at 35℃) with no condensation |
Storage temperature/humidity range |
-20 to 55℃, relative humidity 85% or less(at 35℃) with no condensation |
Size |
79*208*154mm |
Weight |
890g |